Single-patient-use vs disposable vs reusable hoist slings
For occupational therapists, tissue viability and IPC leads, and moving-and-handling teams choosing a sling range.
Three sling categories, three different rules. A disposable sling is issued to one patient for a fixed service life and then binned. A single-patient-use sling is issued to one patient and reused with that patient only, never machine washed. A reusable sling is laundered and reissued between patients under a decontamination policy.
At a glance
| DisposableGo Disposable Slings | Single-patient-useMySling Patient Specific Slings | Reusableemotion Washable and System RoMedic | |
|---|---|---|---|
| Use type | Disposable. The instruction for use is explicit: this is a disposable product and should not be machine washed, recycled or redistributed. | Patient specific. Reused with the same patient only. The instruction for use states it is a patient specific product and should not be machine washed, recycled or redistributed. | Laundered and reissued. The emotion Washable instruction for use states the product may be machine washed in accordance with the label. |
| Typical service life | 6 to 24 weeks, set by transfer frequency, not by appearance: 4 transfers a day gives 6 weeks, 3 a day 8 weeks, 2 a day 12 weeks, 1 a day 24 weeks. Shelf life 4 years boxed, 18 months once unboxed. | Replace after 6 months of use. The date of first use is checked before every transfer. | Service life exceeds 5 years if the laundering guidance is followed. 12-month manufacturer's guarantee. |
| Laundering | None. If the sling is washed or immersed in water the wash label disintegrates to reveal a Do Not Use indicator. | Spot clean only. Safe when wet and safe in a bathroom, shower or toileting environment. A wash indicator changes state if the sling has been machine washed, and it must not be used after that. | emotion Washable: 50 degrees C on a standard cycle for general soiling. For disinfection, a period at temperature of 10 minutes at 65 degrees C and 3 minutes at 72 degrees C. 90 degrees C is possible but moderate shrinkage or colour bleed is likely. |
| Decontamination limits | Not applicable. Dispose of the sling if it is soiled, damaged, or if it is snagged, wrenched or contacts liquids during a transfer. | Not applicable. The sling is not reprocessed. It is replaced. | No bleach, particularly no hypochlorite. No non-biological detergent. No fabric conditioner. Above 100 degrees C causes extreme shrinkage and structural damage. Air dry, or tumble dry cool at 60 degrees C checking every 5 minutes. |
| Attachment | Loop or clip. The loop version fits all 2, 4 and 6 point loop and hook sling bars, subject to risk assessment. | Loop or clip. Purple key-hole clips fit hoist systems with matching fixing points. | Loop or clip depending on the model. Purple key-hole clips on the clip versions. System RoMedic slings can often be used on both 2 and 4 point suspension. |
| Safe working load | 300 kg. The Go Full-Back loop sling is size dependent at 120, 250 or 300 kg. | 300 kg, and 500 kg on XXL sizes. | emotion Washable 300 kg, and 500 kg on XXL sizes. System RoMedic BasicSling 300 kg. |
| LOLER treatment | A sling is lifting equipment used to lift a person, so LOLER 1998 thorough examination at intervals of at most 6 months applies. Use type does not change that. | Same. LOLER applies to a single-patient-use sling exactly as it applies to a laundered one. | Same, and the manufacturer adds an inspection after every laundering. |
| Standards | ISO 10535:2021, with EN ISO 14971:2019 for risk management. Fire performance to EN 1021-1 and EN 1021-2. UKCA and CE marked under Regulation (EU) 2017/745 and the UK Medical Devices Regulations 2002. | ISO 10535:2021, with EN ISO 14971:2019 for risk management. Fire performance to EN 1021-1 and EN 1021-2. UKCA and CE marked under Regulation (EU) 2017/745 and the UK Medical Devices Regulations 2002. Proof tested to 1.5 times the safe working load, held for 20 minutes. | ISO 10535:2021, with EN ISO 14971:2019 for risk management. Fire performance to EN 1021-1 and EN 1021-2. UKCA and CE marked under Regulation (EU) 2017/745 and the UK Medical Devices Regulations 2002. |
| Cost model | Unit cost per patient episode. No laundry cost, no transport, no reprocessing, no downtime waiting on a wash. | One sling per patient for up to 6 months. No laundry cost. The sling is replaced rather than reprocessed. | Lowest unit cost, plus laundry, transport, tracking and downtime. Budget for a spare so the patient is covered while one sling is in the wash. |
| Sizing measurement | Width at hips. | Coccyx to top of head. | Coccyx to top of head and width at hips (emotion). Recommended user weight and colour coded sizes (System RoMedic). |
Which one to choose
Choose disposable when
- The episode is short and the sling will not outlive it.
- The patient is in isolation, or an outbreak makes reprocessing a risk you do not want to carry.
- The sling is going out on community loan and is unlikely to come back.
- There is no laundry capable of holding a disinfection temperature.
- You need the sling today and cannot wait on a wash cycle.
Choose single-patient-use when
- Patient comfort matters. A sling fitted and sized to one person is more comfortable than a disposable sized to a range.
- The stay is long enough for reuse to pay. One sling reused with the same patient replaces several disposables, and the longer the stay the wider the gap.
- One named patient has an ongoing daily need.
- The sling will be used wet, in a bathroom, shower or toileting transfer.
- You want a sling that stays with the patient across settings and is auditable to them by name.
- Laundry capacity is the constraint rather than unit cost.
Choose reusable when
- The setting is stable and has a validated decontamination process.
- The laundry can hold a period at temperature and can evidence it.
- There is enough stock that a patient is never waiting on a wash cycle.
- You need a fabric or a feature only the reusable ranges carry, such as the System RoMedic net materials.
- Unit cost over a multi-year horizon is the deciding factor and the laundry cost is already funded.
What the standards and guidance say
Lisclare's declarations of conformity, issued 2025, apply ISO 10535:2021 as the primary product standard, with EN ISO 14971:2019 for risk management, ISO 20417:2021 and EN ISO 15223-1:2021 for labelling, and EN 1021-1 and EN 1021-2 for fire performance. The slings are CE marked under Regulation (EU) 2017/745 and UKCA marked under the UK Medical Devices Regulations 2002. Lisclare's older Quality and Compatibility Statements cite BS EN 10535:2006, which is superseded. Check the date on any document you are handed.
LOLER 1998 treats a sling as lifting equipment used for lifting a person, which sets a maximum interval of 6 months between thorough examinations. This is a common misunderstanding: a single-patient-use or disposable sling is not exempt because it is not laundered. The duty attaches to the equipment being used to lift a person, not to how the equipment is cleaned.
PUWER 1998 requires that equipment is suitable for the use it is put to and used only by people who have had adequate instruction. Every instruction for use behind this page repeats the same requirement in plain terms: a full risk assessment must be carried out by a suitably qualified individual before any sling is selected.
Both the sling and the hoist carry a safe working load, and every instruction for use makes the same point: the lower of the two limits governs the system. Selecting a 300 kg sling does not give you a 300 kg system if the hoist is rated lower.
Worth noting for an infection prevention and control policy: the point is not only made by the single-patient-use ranges. The System RoMedic manual, which covers laundered multi-patient slings, states plainly that patient-specific slings should be used to ensure safety and hygiene.
What goes wrong
- A single-patient-use or disposable sling goes through the ward laundry. Both ranges are designed to tell you. The MySling wash indicator changes state, and the Go wash label disintegrates to reveal a Do Not Use marker. Once either has happened the sling is finished, and it does not matter how sound the fabric looks.
- A disposable sling is kept in service past its service life because it still looks fine. Service life on the Go range is counted in transfers, not in condition: four transfers a day exhausts a sling in six weeks.
- A reusable sling is put through a 90 degrees C cycle as routine. It is permitted, but shrinkage and colour bleed follow, and on a loop sling the colour is how the correct loop is communicated to the next carer.
- Bleach, and hypochlorite in particular, gets into the wash. It causes structural damage to the sling, and non-biological detergents carry bleaching agents for the same reason.
- A washable sling is shared between patients simply because it is washable, with no decontamination between. Washable describes what the fabric tolerates, not what the infection control policy permits.
- The sling is sized by dress size or by body weight alone. The measurements that matter differ by range: width at hips for Go, coccyx to top of head for MySling, and both for emotion Washable.
- Nobody records the date of first use, so nobody can apply the 6-month replacement point on MySling or the service life on Go. Both instructions for use carry a care plan block for exactly this, with patient name, model, size, issuer and date.
What Lisclare stocks in each column
Generated from Lisclare's product data. Products with an open evidence question are held back rather than listed, so what appears here is only what a named Lisclare document supports.
Disposable
4 products
Single-patient-use
14 products
Reusable
24 products
- AmputeeSling
- Children's HighBack Sling
- Children's HygieneSling
- Children's HygieneVest
- Children's MasterVest
- HighBack Sling
- HygieneSling
- HygieneVest
- LiftPants
- MasterVest Sling
- RepoSheet
- Ultra LiftPants
- UltraSling
- UniversalSling
- AmpSling
- BasicSling
- FlexibleSling
- HighBackSling
- HygieneHBSling
- emotion HighBackSling (clip)
- emotion HighBackSling (loop)
- emotion HygieneSling
- emotion Sit-to-StandSling
- emotion UniversalSling
Questions specifiers ask
- What is the difference between single-patient-use, disposable and reusable slings?
- A disposable sling serves one patient for a set number of transfers and is then binned. A single-patient-use sling serves one patient too, but is reused with that patient for up to 6 months and spot cleaned rather than washed. A reusable sling is laundered to a disinfection temperature and reissued between patients.
- Can single-patient-use slings be washed and reused, or not at all?
- Not at all. MySling slings may be spot cleaned and are safe wet, in a shower or in a toileting transfer, but must not be machine washed. Each one carries a wash indicator that changes state if it has been through a machine, and the instruction for use is explicit: do not use it after that.
- How long does a single-patient-use sling last before it is replaced?
- MySling recommends replacement after 6 months of use, with a visual check before every transfer and the date of first use recorded on the care plan block. Any sign of wear or damage retires it earlier, as does the wash indicator showing it has been machine washed.
- What temperature should a reusable sling be washed at, and is 60 degrees C enough?
- For general soiling the emotion Washable instruction for use specifies 50 degrees C on a standard cycle. For disinfection it specifies a period at temperature: 10 minutes at 65 degrees C or 3 minutes at 72 degrees C. A 60 degrees C cycle is not one of the stated disinfection conditions.
- Do single-patient-use slings still need LOLER inspection?
- Yes. LOLER 1998 applies to lifting equipment used to lift a person, at intervals of at most 6 months, and makes no exemption for slings that are not laundered. In practice the 6-month LOLER interval and the MySling 6-month replacement point fall together, which makes them easy to schedule as one review.
Try before you standardise
Lisclare supplies trial slings and samples so a moving-and-handling team can assess fit, transfer quality and laundry impact on the actual patient and the actual hoist before committing to a range. Trials and samples are covered by the NHS Master Indemnity Agreement with the Department of Health, references IFA822 and IFB822.
Ask for a sling assessment, a sizing visit, or a compatibility check against the hoists already on the ward.
Source documents
- MySling (clip) Instructions for Use, ref SL_MS_C_MI_02
- MySling (loop) Instructions for Use, ref SL_MS_L_MI_03
- MySling Quality and Compatibility Statement, 2025
- MySling Declaration of Conformity, ref SL_MS_DC_02
- Go Disposable Slings (loop attachments) Instructions for Use, ref SL_GL_MI_01
- Go Disposable Slings (clip attachments) Instructions for Use, ref SL_GC_MI_05
- Go Disposable Sling Declaration of Conformity, ref SL_G_DC_03
- emotion Washable UniversalSling Instructions for Use, ref SL_EWUS_MI_01
- emotion Sling Declaration of Conformity, ref SL_EM_DC_02
- SystemRoMedic BasicSling manual, no. 00848 En Rev 7
Last reviewed 2026-09-07. Draft for Lisclare review. Not signed off.